What should I do if the constant temperature and humidity chamber is in a water shortage state?

When the constant temperature and humidity chamber is used for a long time, the equipment has a water shortage alarm, an overpressure alarm, and an overload alarm. These problems are not caused by equipment failures, and can be solved by related manual operations such as resetting. Therefore, users do not have to worry too much. This method is simple and fast. The most important thing is that it will not affect the experiment. However, in the use of the equipment, there will also be water droplets in the upper right corner of the studio. This problem has a great impact on the test, so the user should pay special attention to the countermeasures.

First, you need to find out the cause of the problem and then prescribe the right medicine. Open the studio door in the upper right corner will see there are so many parts, over temperature protector, temperature sensor, humidity sensor, wet bulb, the normal order should also be arranged in this way, but some more powerful manufacturers The over-temperature protector and the temperature sensor are combined on one sensor, and only the temperature sensor, the humidity sensor, and the wet bulb are in the working chamber, and the humidity gauze is hung on the humidity sensor in a suspended state, and the bottom end is directly Dip in the wet bulb, if there is a drop of water, it can be observed by opening the door.

Secondly, a pair of waterways are inspected. It is necessary to observe whether the inlet valve is too large, resulting in too much water. If this is the case, the water valve can be directly closed. 2 Check if the hydration time is set too high to avoid excessive water supply. If this is the case, the hydration interval can be increased to reduce the frequency. 3 Check whether the floor of the laboratory is level. The angle of the tilt will also cause the water of the wet bulb to be unbalanced, so that the water will drip down. At this time, just move the equipment to the flat ground and use it. 4 Check whether the water level box on the right side of the constant temperature and humidity chamber is too high to prevent the water from entering too much. At this time, the water level box needs to be adjusted. For this situation, it is recommended to adjust the water level box with professional technology. The engineer conducts professional communication and then processes to avoid the adverse consequences of the wrong action.

Finally, if the above content can not help you solve the problem, you can consult Shanghai Linpin Instrument Co., Ltd., which has professional engineers to serve you online!

What should I do if the constant temperature and humidity chamber is in a water shortage state?

Shoulder Bag

PU Shoulder Bag has many designs to choose, it's very convenient to carry, which is more handy than backpacks. It can pack as many things as possible in the smallest space, which is very popular among students and office workers. , can also be used as a Storage Bag, is a multi-purpose storage supplies.756

Shoulder Bag,Luminous PU Shoulder Bag,Cork Shoulder Bag,Foldable Shoulder Bag,Fashion Shoulder Bag,Geometric Shoulder Bag.

Jilin Y.F. Import & Export Co.,Ltd , https://www.yfpencilcase.com