Analysis of slow temperature drop in constant temperature and humidity test chamber

Analysis of slow temperature drop in constant temperature and humidity test chamber

Because customers often encounter the situation that the constant temperature and humidity equipment is not cooled, so today we will study and discuss this issue together:
Reason one:
1. Because the temperature cannot be maintained, observe whether the refrigeration compressor can be started during the operation of the test box and the compressor can be started during the operation of the environmental test equipment, indicating that the electrical circuit from the main power supply to each compressor is normal. There is no problem with the system.
2. There is no problem with the electrical system, continue to check the refrigeration system. First check the discharge and suction pressures of the low-temperature (R23) compressors of the two groups of refrigeration units are lower than normal, and the suction pressure is evacuated, indicating that the main refrigeration unit has insufficient refrigerant quantity.
3. Touch the exhaust and suction lines of the main unit R23 compressor by hand, and find that the temperature of the exhaust line is not high, and the temperature of the suction line is not low (no frost), which also shows the main unit The lack of R23 refrigerant.
Reason two:
1. The cause of the failure is not determined, and the cause of the failure is further confirmed in conjunction with the control process of the test box. The test box has two sets of refrigeration units.
2. One is the main unit and the other is the auxiliary unit. When the cooling rate is large, the two groups of units work at the same time. At the beginning of the temperature maintenance stage, the two groups of units still work at the same time. After the temperature stabilizes initially, the auxiliary unit stops working, and the host group maintains the temperature stability. If the host group R23 leaks, the cooling effect of the host group will be insignificant. Since the two units are working at the same time during the cooling process, there is no phenomenon that the temperature is not stable, and the cooling rate is indicated to decrease. In the temperature maintenance stage, once the auxiliary unit stops working and the main unit has no cooling effect, the air in the test chamber will slowly rise. When the temperature rises to a certain level, the control system will start the auxiliary unit to cool down and lower the temperature to the set point. Near the fixed value (-55 ℃), then the auxiliary unit stops working again, and the fault phenomenon as shown in Fig. 3 will appear after so repeated.
So far, it has been confirmed that the cause of the production failure is the leakage of refrigerant R23 of the low-temperature (R23) class unit of the host group. The leak detection of the refrigeration system was checked by a combination of a leak detector and soapy water, and it was found that the stem of a hot gas bypass solenoid valve was cracked by a slit of about 1 cm. Replace the solenoid valve, recharge the system with fluorine, and the system is operating normally. As can be seen from the above, the analysis and judgment of this fault phenomenon are basically easy to difficult. The analysis and judgment of the "outer" then "in", "electrical" and then "cooling" are familiar and familiar. Understanding the principle and working process of the test chamber is the basis for analyzing the fault and judging the fault.

Orlistat is a long-acting and potent specific gastrointestinal lipase inhibitor. It is white or off-white powder at room temperature. It is insoluble in water, soluble in chloroform, and easily soluble in ethanol. It passes through the stomach and small intestine cavity. Intragastric lipase and pancreatic lipase active codon form a covalent bond to inactivate the enzyme. Fat in food cannot be broken down into free fatty acids and monoacylglycerols, so the fat cannot be absorbed and utilized, thereby reducing the body's caloric intake and controlling body weight. This medicine does not need to be absorbed through the body to exert its effects. At the usual dose, fat absorption can be suppressed by 30%. It is rarely absorbed after oral administration and can be metabolized and inactivated in the intestine. The metabolic site is on the wall of the gastrointestinal tract, and the elimination half-life is about 14 to 19 hours. About 97% of this product is excreted with feces, of which 83% is excreted in the original form.

Orlistat can be used clinically for obesity and hyperlipidemia. Under normal circumstances, 120mg can be taken orally once a day, three times a day, and taken within one hour after a meal. After taking the medicine for 2 weeks, the weight may start to decrease. It can be taken continuously for 6 to 12 months. If the dose is increased to more than 400 mg per day, its effect will no longer be enhanced.

Orlistat

Orlistat Dosage,Orlistat Weight Loss,Orlistat Function,Orlistat For Weight Loss

Jinan Prosperity International Trade Co. Ltd , https://www.jnprosperity.com